← All lots
Lot verification
GLP-3 RT 20 mg
Lot GLP3RT-20-P8Elab lot ckxfpz9
This lot is currently available.
Chromatographic purity
By HPLC-UV/MS · our standard: at least 98%
99.9%
Meets standard
Content vs. label
19 mg measured against a 20 mg label · accepted range 90–110%
94.4%
Meets standard
Identity
Confirmed by HPLC-UV/MS
Confirmed
Meets standard
Tested by Krause Analytical
Testing commissioned by Finnrick, not by Amino+ Labs.
Analysed 2026-05-15 · reported 2026-05-15
The certificate
Lot pages are permanent. Sold-out and superseded lots keep their results published so this archive stays checkable.
For research use only. Not intended for human or veterinary use. Not approved by the FDA for clinical use.